ClearanceBrief
Recall Z-2651-2014 ยท GIF Diluent, Blood Cell

Beckman Coulter Inc.

Coulter DxH Diluent, Catalog No. 628017 Blood cell analysis comprises diluting a whole-blood sample with a solution that functions as a diluent. The diluent provides the ability to analyze portions of the diluted blood sample for different blood cell types and the measurement of hemoglobin.

DateSep 16, 2014 (initiated Aug 25, 2014)
StatusTerminated
ReasonBeckman Coulter is recalling Coulter DxH Diluent because it may be associated with elevated platelet (PLT) background counts between 3-7 x 10(3) cells/uL. This will result in failed Daily Checks on the UniCel DxH 800 and DxH 600 Coulter Cellular Analysis Systems.
Root causeUnder Investigation by firm
ActionBeckman Coulter sent an Urgent Medical Device Recall letter dated August 25, 2014, to all customers that purchased the Coulter DxH Diluent. The letter informed the customers of the problems identified and the actions to be taken. Customers were instructed to complete and return the enclosed response form within 10 days. Customers with questions were instructed to contact Beckman Coulter's Customer Support Center at http://www.beckmancoulter.com, or call (800) 526-7694 in the US and Canada.
Quantity46,050 units total (40,078 units in US)
DistributionWorldwide Distribution - USA (nationwide) and Internationally to Canada, Chile, Mexico, Panama, and Thailand. .
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.