Recall Z-2685-2014 ยท PCH Gastrointestinal Pathogen Panel Multiplex Nucleic Acid-Based Assay System
Nanosphere, Inc.
The Verigene Enteric Pathogens Nucleic Acid Test (EP) is a multiplexed, qualitative test for simultaneous detection and identification of common pathogenic enteric bacteria and genetic virulence markers from liquid or soft stool preserved in Cary-Blair media, collected from individuals with signs and symptoms of gastrointestinal infection. The test is performed on the automated Nanosphere Verigene
| Date | Sep 23, 2014 (initiated Oct 10, 2013) |
| Status | Terminated |
| Reason | Nanosphere has received several reports relating to an increased rate of Processing Errors during procedures performed on the Processor SP caused by Tips in Tip Holder Assemblies for the following products and lots of Extraction Trays. |
| Root cause | Nonconforming Material/Component |
| Action | Nanosphere sent a Field Notification Letter dated October 10, 2013, to all affected customers. The letter identified the product, the problem and the action to be taken by the customer. Instructions were included in the letter for customers to determine whether or not the extraction tray lots should be returned. If the extraction tray lots are to be returned, customers were to complete and return the attached form and replacement tip holder assemblies will be supplied at no cost. Customers could then dispose of the recalled unused tip holder assemblies. Customers with questions were instruc |
| Quantity | 220 units |
| Distribution | Worldwide Distribution - USA including AZ, CA, FL, GA, IA, IL, IN, KY, MA, MD, MN, MO, MT, NC, NE, NJ, NY, OH, OR, PA, TX, UT, VA, WA, WI, OH and Internationally to Bulgaria, Germany, Italy, Japan, South Korea, and Spain. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K140083 | VERIGENE ENTERIC PATHOGEN NUCLEIC ACID TEST ( EP) Nanosphere, Inc., Jun 20, 2014 |