Recall Z-2678-2014 ยท JAX Pneumotachometer

Philips Medical Systems (Cleveland) Inc

No packaging. The device is a computed tomography scanner. Intended to provide corss sectional images of the human body and visualization of the internal organs of the body.

DateSep 22, 2014 (initiated Jan 9, 2007)
StatusTerminated
ReasonWhen the Extended Display Firled ov View reconstruction is used, the CT numbers and anatomy in teh extended region are not accurate.
Root causeLabeling design
ActionPhilips sent a Product Safety Notification letter dated December 21, 2006 to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to distribute the Product Safety Notification to all affected users and keep a copy of the notification with the Instructions for Use until further notice. Customers with questions were instructed to contact Customer Care Center at 1-800-722-9377, option 5 - Diagnostic Imaging, option 1 - CT. For questions regarding this recall call 440-483-7600.
Quantity61
DistributionWorldwide Distribution - US including MO, TX, MD, AR, VA, KY, RI, NJ, MA, NY, LA, PA, VT, SC, CA, OH, FL, MN, MI and Internationally to Netherlands, Canada, Finland, Austria, France, Malaysia, Denmark, Norway, UK, Australia, Belgium, New Zealand.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K033357ACQSIM-MULTISLICE-CT, MODEL 453567042271
Philips Medical Systems (Cleveland), Inc., Jan 9, 2004