ClearanceBrief
Recall Z-0026-2015 ยท LJS Catheter, Intravascular, Therapeutic, Long-Term Greater Than 30 Days

Navilyst Medical, Inc

BioFlo PICC with ENDEXO Technology and PASV Valve Technology Maximal Barrier Nursing Kit, 5F-55 cm, UPN Product No. H965750191, REF/Catalog No. 75-019, STERILE --- Indicated for short or long-term peripheral access to the central venous system for intravenous therapy, including but not limited to, the administration of fluids, medications, nutrients; the sampling of blood and for power injection

DateOct 7, 2014 (initiated Aug 25, 2014)
StatusTerminated
ReasonNMI was notified by the manufacturer of the 10 ml BD PosiFlush SF Saline Flush Syringes, that it had initiated a recall on specific lots of the syringes as some syringe unit packages exhibited open seals which may impact package integrity and potentially affect product sterility. Based on the notification, NMI is recalling specific PICC Kits that contain the affected 10 ml BD PosiFlush SF Saline Flush Syringes.
Root causeNonconforming Material/Component
ActionBD Medical sent an Urgent Voluntary Medical Device Recall letter dated July 22, 2014, to all affected customers via Federal Express on August 25, 2014. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to segregate and quarantine all affected product. Immediately discontinue shipment of the affected product. Complete the Recall Response Card form and fax it back to BD at 1-201-847-6990. If the affected product was further distributed cusotmers should identify their cusotmers and notify them immediately of the product reca
Quantity300 units/pouches
DistributionNationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K0420610.9% SODIUM CHLORIDE INJECTION, USP BD POSTFLUSH SF FLUSH SYRINGE
Becton Dickinson, Aug 26, 2004
K121990PICC CONVENIENCE KIT
Navilyst Medical, Inc., Aug 3, 2012
K131038NMI PICC MAXIMAL BARRIER NURSING KITS
Navilyst Medical, Inc., May 9, 2013