ClearanceBrief
Recall Z-0975-2015 ยท JSG Culture Media, Non-Selective And Non-Differential

Becton Dickinson & Co.

CDC Anaerobe 5% Sheep Blood Agar CDC Anaerobe 5% Sheep Blood Agar is an enriched, nonselective culture medium particularly useful or the isolation and cultivation of obligate anaerobes from clinical specimens. It supports the growth of a wide variety of obligately anaerobic, faculatively anaerobic, microaerophilic and aerobic bacteria.

DateJan 14, 2015 (initiated Jun 6, 2014)
StatusTerminated
ReasonA low level of surface and subsurface contamination of Listeria monocytogenes on 9 lots of non-sterile prepared plated media containing sheep blood, typically visible immediately upon removal from the packaging. The contamination is typically observed as 1-3 CFUs per plate when present.
Root causeMaterial/Component Contamination
ActionBD Diagnostic Systems sent a Field Correction Notice dated August 2014 to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to continue to follow all surveillance requirements of their laboratory upon confirmation of Listeria monocytogenes. Laboratory staff should notify their respective state health department. Customers with questions should contact BD Technical Services at 800-638-8663. For questions regarding this recall call 410-316-4000.
Quantity1,530 cases (100 plates/case)
DistributionWorldwide Distribution - US nationwide including Puerto Rico, Belgium, Canada, Philippines, Singapore, and Taiwan.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K803025CDC ANAEROBE BLOOD AGAR
Bd Becton Dickinson Vacutainer Systems Preanalytic, Dec 22, 1980