ClearanceBrief
Recall Z-0364-2016 ยท KPR System, X-Ray, Stationary

Philips Medical Systems, Inc.

Patient Support Material Number: 9890 010 87431 used with: BuckyDiagnost ; Model: 704060/706037

DateAug 19, 2014
StatusTerminated
ReasonThe hook does not securely hold the footplate in vertical position. Therefore the footplate can fall or swing down and may hurt the operator or patient.
Root causeDevice Design
ActionThe Field Safety Notice FCO70600086 will be mailed on August 19, 2014. Philips Medical Systems will send out certified letters and will have the receipts returned for tracking purposes as well as have a Field Service Engineer visit each site to provide the fix when the Field Change Order is released.
Quantity1517 systems, not all units have the patient support accessory
DistributionDistributed in the states of MN, KY, MA, VA, TX, NC, HI, FL, AZ, CT, and WY and the countries of Argentina, Australia, Austria, Canada, Denmark, Germany, Netherlands, Russia, Saudi Arabia, Singapore, Spain, and Switzerland.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K945278PHILIPS BUCKY DIAGNOST FAMILY
Philips Medical Systems North America, Inc., Dec 14, 1994