ClearanceBrief
Recall Z-2572-2014 ยท LWS Implantable Cardioverter Defibrillator (Non-Crt)

St Jude Medical Cardiac Rhythm Management Division

Ellipse Implantable Cardioverter Defibrillators (ICDs), Model No. CD1275, CD1311, CD2275, CD2311, CD1411, CD2411 Ellipse ST Implantable Cardioverter Defibrillators (ICDs), Model No. CD1273, CD1309, CD1409, CD2273, CD2309, CD2409.

DateSep 3, 2014 (initiated Aug 19, 2014)
StatusTerminated
ReasonAn anomaly may result in delayed, partial, or no delivery of high voltage therapy. Internal damage to the high voltage capacitors, which may result in extended charge time, occurring during capacitor maintenance or charging for high voltage therapy. It may present as a vibratory patient notification and upon interrogation, a programmer or Merlin.net alert.
Root causeComponent design/selection
ActionA customer letter dated 8/19/14 was sent to physicians with the affected Ellipse ICDs on 8/20/14. The letter identifies the problems and the actions to be taken. Customers with any questions or concerns are instructed to contact their local St. Jude Medical representative or St. Jude Medicals Technical Services Department.
Quantity51,492 units (27,396 units in US)
DistributionWorldwide Distribution: US (nationwide), Canada, Australia, New Zealand, Japan, EU, Middle East, Latin America, and Asia-Pacific.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.