ClearanceBrief
Recall Z-2650-2014 ยท KFF Assay, Heparin

Diagnostica Stago, Inc.

Stago IVD, Liquid Anti-Xa, NDC 3607450003116. The STA - Liquid Anti-Xa kits are intended for use with STA-R, STA Compact and STA Satellite Analyzers for the quantitative determination of the plasma levels of unfractionated (UFH) and low molecular weight (LMWH) heparins by measuring their anti-Xa activity on antithrombin in a competitive assay using a synthetic chromogenic substrate.

DateSep 16, 2014 (initiated Jul 18, 2014)
StatusTerminated
ReasonDiagnostica Stago Inc. received some customer complaints regarding a misreading of Reagent 1 (substrate) barcode on STA analyzers.
Root causeUnder Investigation by firm
ActionDiagnostica Stago sent a "Quality Information" letter/Customer Acknowledgement form to its customers. The letter identified the product, problem, and actions to be taken by the customers. For additional information and further support contact the Diagnostica Stago Hotline at 800-725-0607.
Quantity514 kits
DistributionDistributed USA (nationwide) and the country of Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K111822STA(R) - LIQUID ANTI-XA, MULTI HEP CALIBRATOR, QUALITY HNF / UFH, QUALITY HBPM / LMWH
Diagnostica Stago, Inc., Oct 26, 2011