Recall Z-2491-2014 ยท LRO General Surgery Tray

Customed, Inc

Breast Pack MRT, code 900-2658, contains: ( 1) BULB SYRINGE 60cc (2) GOWN SOFT SMS STD XLGE SET IN SLEEVE ( 1) CAUTERY PENCIL PUSH BOTTOM (2) SYRINGE 10cc W/0 NEEDLE LUER LOCK ( 1) NEEDLE and BLADE COUNTER 20c FOAM/MAG ( 1) UTILITY BOWL QUART 32oz ( 1) MAYO STAND COVER REINFORCED ( 1) TABLE COVER REINFORCED 50 X 90 ( 1) DRAPE BREAST/CHEST SMS ( 1) YANKAUER SUCTION TUBE W/0 VENT ( 1) TUBE SUCTION C

DateAug 29, 2014 (initiated May 20, 2014)
StatusTerminated
ReasonCustomed has determined that there is the possibility that packaging integrity may be compromised on different catalog numbers (multiple lots). This could result in an injury to the patient due to product contamination or loss of sterility condition.
Root causePackage design/selection
ActionCustomed Inc. sent a Recall Notification letter dated June 3, 2014 via email and followed by letter to all affected customers. The letter identified the affected products, problem and actions to be taken. The letter informed customers if the affected products are in their inventory to: 1. Remove products from sale (quarantine them) and cease distribution. 2. Complete and return the enclosed acknowledgement form to Customed Inc .Recall Coordinator. Affected products should be returned to Customed Inc for credit. Customers with questions were instructed to contact Customed Inc. Recall Coor
Quantity3 lots, 306 units, multiple units per lot
DistributionWorldwide Distribution - US (nationwide) in the state of Florida including Puerto Rico (USA); & USVI (St John) and the countries of Tortola, Trinidad, Dominican Republic, Costa Rica and Mexico.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.