Recall Z-0009-2015 ยท EHD Unit, X-Ray, Extraoral With Timer

Gendex Corp

eXpertDC intraoral x-ray, models: eXpert DC, NA, 75 Inch reach; eXpert DC, NA, 65 Inch reach; eXpert DC, CN, 55 Inch reach; eXpert DC, WW, 75 Inch reach; eXpert DC, WW, 65 Inch reach; eXpert DC, WW, 55 Inch reach; OEM Arm Assembly w/o cables; Assy, Articulated Arm & Tubehead, eDC; and Assy, Articulated Arm & Tubehead, 765DC Product Usage: intraoral x-ray

DateOct 2, 2014 (initiated Aug 12, 2014)
StatusTerminated
ReasonTwo (2) failures of the spring link components within the articulated arm assembly. The spring links broke during normal assembly process for the articulated arm during first cycle of the arm, prior to the 14lb tube head (full load) being installed onto the assembled articulated arm.
Root causeNonconforming Material/Component
ActionGendex sent a Recall Notification Letter, dated August 12, 2014 and a second Communication Notification letter on September 22, 2014 to all affected customers. The letter identified the affected product, problem, and actions to be taken. For questions contact Gendex Customer Service at 1-800-323-8029.
Quantity651
DistributionWorldwide Distribution - US Nationwide and the countries of Europe, Canada, Australia, China, Russia, Taiwan, Singapore, and Thailand.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K992610GENDEX 765DC, MODEL 110-0154
Dentsply Intl., Sep 7, 1999