Recall Z-2628-2014 ยท JWY Manual Antimicrobial Susceptibility Test Systems
Becton Dickinson & Co.
BD BBL Vancomycin Screen Agar Catalog number 222204 Vancomycin Screen Agar is used to test enterococci for resistance to vancomycin and to predict the synergistic activity of this antimicrobial with an aminoglycoside antimicrobial.
| Date | Sep 12, 2014 (initiated Jul 28, 2014) |
| Status | Terminated |
| Reason | These lots of Agar may have exhibited breakthrough growth of vancomycin susceptible Enterococcus species control organisms. Agar that allows susceptible enterococci to grow appears as a falsely resistant culture. If a lab does not QC, the error would be identified as falsely resistant enterococci would have further AST workup. There could be a delay in the identification of false-resistance |
| Root cause | Under Investigation by firm |
| Action | BD Diagnostic System sent an Urgent Product Recall letter dated July 2014, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Our records indicate you may have been shipped the above-referenced lot numbers of product between March 5, 2014 and June 5, 2014. Please discontinue use of these lot numbers and discard any remaining packages. If further assistance is needed, please contact BD Technical Services Department at 1-800-638-8663. The acknowledgement form included with the follow up letter that is faxed back to BD wil |
| Quantity | 28540 |
| Distribution | Worldwide Distribution: US (Nationwide) including the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, VT, WA, WI and WV; and Internationally to the countries |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K951539 | ENTEROCOCCUS SCREENING MEDIUM Becton Dickinson Microbiology Systems, Jun 13, 1995 |