ClearanceBrief
Recall Z-2482-2014 ยท OFA Cardiovascular Surgical Instruments Tray

Customed, Inc

Open Heart Pack, code 900-2618, contains: (9) DRAPE 60" X 76" REINFORCED LIF ( 1) MAYO STAND COVER REINFORCED LIF (6) TOWELS ABSORBENT 15" X 20" L/F ( 1) TABLE COVER BTC 77" X 110" HD FULL COVERAGE LIF ( 1) MAYO TRAY LARGE (8) DRAPE UTILITY WfTAPE LIF (1 ) TABLE COVER REINFORCED 44" X 78" LIF (2) GOWN SURG. SMS POLY-REINF BREATHABLE AAMI IV X-LARGE T/WRAP ( 1) DRAPE SPLIT CARDIO W/CLEAR ANEST. SC

DateAug 29, 2014 (initiated May 20, 2014)
StatusTerminated
ReasonCustomed has determined that there is the possibility that packaging integrity may be compromised on different catalog numbers (multiple lots). This could result in an injury to the patient due to product contamination or loss of sterility condition.
Root causePackage design/selection
ActionCustomed Inc. sent a Recall Notification letter dated June 3, 2014 via email and followed by letter to all affected customers. The letter identified the affected products, problem and actions to be taken. The letter informed customers if the affected products are in their inventory to: 1. Remove products from sale (quarantine them) and cease distribution. 2. Complete and return the enclosed acknowledgement form to Customed Inc .Recall Coordinator. Affected products should be returned to Customed Inc for credit. Customers with questions were instructed to contact Customed Inc. Recall Coor
Quantity7 lots, 259 packed units, multiple units for lot.
DistributionWorldwide Distribution - US (nationwide) in the state of Florida including Puerto Rico (USA); & USVI (St John) and the countries of Tortola, Trinidad, Dominican Republic, Costa Rica and Mexico.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.