ClearanceBrief
Recall Z-2456-2014 ยท LRO General Surgery Tray

Customed, Inc

EXTREMITY ARM PACK CONTENTS: (1) STOCKINETTE IMPERVIOUS 14" X 48" (1) GOBAN SELF-ADHERENT WRAP 6" X 5yds (3) BANDAGE ELASTIC 4" X 5yds (1) BANDAGE ESMARK 6" X 9 (1) UTILITY BOWL QUART 32oz (10) GAUZE SPONGE 4" X 4" 16PLY (1) TUBE SUCTION CONNECTX 12' (1) SKIN MARKER INK W/8 LABEL/TIME OUT/RULER (3) SCALPEL WITH HANDLE# 15 STAINLESS STEEL (1) TIME OUT BEACON NON WOVEN (1) LIGHT

DateAug 29, 2014 (initiated May 20, 2014)
StatusTerminated
ReasonCustomed has determined that there is the possibility that packaging integrity may be compromised on different catalog numbers (multiple lots). This could result in an injury to the patient due to product contamination or loss of sterility condition.
Root causePackage design/selection
ActionCustomed Inc. sent a Recall Notification letter dated June 3, 2014 via email and followed by letter to all affected customers. The letter identified the affected products, problem and actions to be taken. The letter informed customers if the affected products are in their inventory to: 1. Remove products from sale (quarantine them) and cease distribution. 2. Complete and return the enclosed acknowledgement form to Customed Inc .Recall Coordinator. Affected products should be returned to Customed Inc for credit. Customers with questions were instructed to contact Customed Inc. Recall Coor
Quantity4 lots/112 units (multiple units per lot)
DistributionWorldwide Distribution - US (nationwide) in the state of Florida including Puerto Rico (USA); & USVI (St John) and the countries of Tortola, Trinidad, Dominican Republic, Costa Rica and Mexico.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.