Recall Z-2719-2014 ยท GZB Stimulator, Spinal-Cord, Implanted (Pain Relief)

St. Jude Medical, Inc.

Percutaneous Trial Lead Kit Model 3086 (8 ch percutaneous lead). Used as part of the St. Jude Medical Neurostimulation system indicated for Spinal Chord Stimulation.

DateSep 29, 2014 (initiated Jul 29, 2014)
StatusTerminated
ReasonThe Directions for Use (DFU) manual contains information for devices that are not approved in the US. The correct US approved versions of the manuals were sent to the customers to replace the incorrect international manuals.
Root causePackaging process control
ActionA field action was initiated with customer letters, dated July 29, 2014, and attached replacement DFUs. The letter identified the affected product and the reason for the recall. The recalling firm requested the consignees refer to the replacement DFU for guidance while using the device, and to discard the international manuals. Questions regarding the letter should be directed to customers' St. Jude Neuromodulation Representative or call the Technical Services Department at 1-800-727-7846.
Quantity4 units
DistributionDistributed in the states of Texas and Illinois.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K123299MTS SYSTEM, WIDE-SPACE QUATTRODE PERCUTANEOUS LEADS
St. Jude Medical Neuromodulation, Jan 18, 2013