Recall Z-2206-2014 ยท FLD Apparatus, Hemoperfusion, Sorbent

Gambro Renal Products, Incorporated

PrisMARS Kit (component of MARS Treatment Kit Type 1115/1 PrisMARS US), 1 MARS Treatment Kit per carton box, Product Number 800480, Batch

DateAug 13, 2014 (initiated Jul 18, 2014)
StatusTerminated
ReasonGambro initiated a field action on the Prisma and Prismaflex sets due to breakages of connectors resulting in leakage of liquid or blood.
Root causeDevice Design
ActionGambro sent an Urgent Field Safety Notice dated July 18, 2014 to all end-users. A Customer Reply Form was attached to the notice for customers to complete and return.
Quantity401,451
DistributionWorldwide Distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.