ClearanceBrief
Recall Z-2578-2014 ยท JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer

Zimmer, Inc.

PERSONA The Personalized Knee System Tibial Articular Surface Provisional Left with the following: Size CD +0 mm Thickness Bottom(TB); Size CD +6 mm Thickness Bottom; Size EF +0 mm Thickness Bottom; Size EF +6 mm TB; Size GH +0 mm TB; Size GH +6 mm TB; Size J +0 mm TB; Size J +6 mm TB; Right Size 1-2 CD +0 mm TB; Right Size CD +6 mm TB; Right Size EF +0 mm TB; Right Size EF +6 mm TB; Right Size

DateSep 4, 2014 (initiated Aug 7, 2014)
StatusTerminated
ReasonReports of breakage of the Persona Knee Trial Articular Surface Provisionals (TASPs). The Persona primary knee surgical technique has been updated. This field action correction is to ensure that all distributors and users of this instrument are aware of the enhanced surgical techniques.
Root causeDevice Design
ActionOn 8/7/2014, Zimmer notified all distributors via electronic mail. Distributors that have received affected inventory are also notified via courier. Hospital risk managers and surgeons are also notified via courier. Hospital risk managers and surgeons were provided with a notification identifying the issue and their responsibilities. These responsibilities include ensuring the affected personnel are aware of the contents of the notification. Distributors were sent a notification identifying the issue and to provide a copy of the notification sent to surgeons and risk managers. The distrib
Quantity77,737 total units
DistributionWorldwide Distribution: US (nationwide) including states of: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, and WY; and countries of: AUSTRALIA, INDIA,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K113369ZIMMER PERSONA KNEE SYSTEM
Zimmer, Inc., Mar 27, 2012
K123459ZIMMER PERSONA TM PERSONALIZED KNEE SYSTEM
Zimmer, Inc., Feb 22, 2013