Recall Z-2577-2014 ยท JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer
Zimmer, Inc.
PERSONA The Personalized Knee System CR Tibial Articular Surface Provisional with the following: Left Size 3-9 CD Top; Left Size 3-11 EF Top; Left Size 7-12 GH Top; Left Size 9-12 J Top; Left Size 3-9 CD Top; Left Size 3-11 EF Top; Right Size 7-12 GH Top; and Right Size 9-12 J Top.
| Date | Sep 4, 2014 (initiated Aug 7, 2014) |
| Status | Terminated |
| Reason | Reports of breakage of the Persona Knee Trial Articular Surface Provisionals (TASPs). The Persona primary knee surgical technique has been updated. This field action correction is to ensure that all distributors and users of this instrument are aware of the enhanced surgical techniques. |
| Root cause | Device Design |
| Action | On 8/7/2014, Zimmer notified all distributors via electronic mail. Distributors that have received affected inventory are also notified via courier. Hospital risk managers and surgeons are also notified via courier. Hospital risk managers and surgeons were provided with a notification identifying the issue and their responsibilities. These responsibilities include ensuring the affected personnel are aware of the contents of the notification. Distributors were sent a notification identifying the issue and to provide a copy of the notification sent to surgeons and risk managers. The distrib |
| Quantity | 77,737 total units |
| Distribution | Worldwide Distribution: US (nationwide) including states of: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, and WY; and countries of: AUSTRALIA, INDIA, |
| FDA record | Open on fda.gov |