Recall Z-2671-2014 ยท MZO Assay, Enzyme Linked Immunosorbent, Hepatitis C Virus

ORTHO-CLINICAL DIAGNOSTICS

VITROS Immunodiagnostic Products Anti-HCV Reagent Pack for in vitro qualitative detection of immunoglobulin G antibody to hepatitis C virus (anti-HCV) in human serum and plasma. REF 680 2454, IVD -- Ortho-Clinical Diagnostics High Wycombe UK. Note that one sales unit has 5 Reagent Packs contained within it. Each Reagent Pack can test for 100 samples (patient, Quality Control, etc.).

DateSep 19, 2014 (initiated Aug 6, 2014)
StatusTerminated
ReasonVITROS Immunodiagnostic Products Anti-HCV kit lot 9090 has been observed producing sporadic lower than expected VITROS Anti-HCV test results due to an approximate signal loss of up to 66.2% which can result in unexpected negative test results being obtained for reactive sample fluids (quality control and patient samples).
Root causePackaging process control
ActionAll US consignees (users) who were shipped the affected lot VITROS HCV Reagent Pack were sent communications via FedEx and/or US Priority Mail (PO Boxes only) on 8/06/14, and advised of the issue and to immediately discontinue using all remaining inventory of VITROS Anti-HCV Reagent Packs Lot 9090. For additional information, please contact OCD's Customer Technical Service representatives at 1-800-421-3311.
Quantity463 sales units
DistributionNY, NJ, CT, MA, PA, VA, MD, FL and GA
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.