Recall Z-2561-2014 ยท JQP Calculator/Data Processing Module, For Clinical Use

Siemens Healthcare Diagnostics, Inc.

Siemens EasyLink Informatics System V4.0/V5 Siemens, Material Number (SMN) 10444799, 10444800, 10487219 A clinical data management system that assists medical laboratory professionals with preanalytic and post-analytic functions in conjunction with multiple instruments, the laboratory information system (LIS) and Siemens StreamLAB Automation Solutions.

DateSep 1, 2014 (initiated Aug 13, 2013)
StatusTerminated
ReasonMay not perform as intended under certain conditions, causing the release of results to the Laboratory Information System (LIS) that should have been held for manual review due to auto-verification rules or the delay/omission of result transmission to the LIS.
Root causeSoftware design
ActionSiemens issued an Urgent Medical Device Correction (UMDC) letter on August 13, 2013 to all affected customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to review this letter in its entirety and ensure that all operators understand the information presented. The issues described are specific to the features listed in the letter.. If the laboratory is not using a particular feature, then they are not affected by the issues described for that feature. For questions contact your Siemens Customer Care Center or your local Siemens techn
Quantity1347
DistributionUS Nationwide Distribution