Recall Z-2300-2014 ยท JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer
Zimmer, Inc.
Persona (TM) The Personalized Knee System Tibial Articular Surface Provisional Bottom Nonsterile PSN A/S PROV L CD BOTTOM +0MM PSN A/S PROV L CD BOTTOM +6MM PSN A/S PROV L EF BOTTOM +0MM PSN A/S PROV L EF BOTTOM +6MM PSN A/S PROV L GH BOTTOM +0MM PSN A/S PROV L GH BOTTOM +6MM PSN A/S PROV L J BOTTOM +0MM PSN A/S PROV L J BOTTOM +6MM PSN A/S PROV R CD BOTTOM +0MM PSN A/S PROV R CD BOTTOM +6MM PSN
| Date | Aug 21, 2014 (initiated Jun 10, 2013) |
| Status | Terminated |
| Reason | Zimmer had received complaints that reported breakage of the Persona Knee Trial Articular Surface Provisionals (TASPs). This is a retrospective recall which covers Zimmer's corrective actions taken earlier this year concerning the TASPs. |
| Root cause | Device Design |
| Action | On February 10, 2014, Zimmer sent letter to surgeons who had filed complaints regarding the affected TASP fractures. The letter provided the surgeons with Zimmer's investigational findings of the issue. Surgeons were directed to send questions or any additional informtion to Zimmer.PER@Zimmer.com. The Zimmer Weekly Wrap-Up, dated February 14, 2014, was issued electronically to sales force. They were notified to engage in mandatory training to improve the usage of the TASPs and prevent potential breakage. Zimmer has taken additional corrective measures by initiating another recall in July 2 |
| Quantity | 70,986 distribution events |
| Distribution | Worldwide Distribution-USA (nationwide) including DC and Puerto Rico and the states of AK, AL, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, MA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, SD, TN, TX, UT, VA, WA, WI, WY, and the countries of Australia, Canada, Germ |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K123459 | ZIMMER PERSONA TM PERSONALIZED KNEE SYSTEM Zimmer, Inc., Feb 22, 2013 |