Recall Z-2551-2014 ยท JDG Prosthesis, Hip, Femoral Component, Cemented, Metal
Zimmer, Inc.
Moore & Thompson Hip Stems: FNSTRD HIP 38HDX127ST FNSTRD HIP 40HDX127ST FNSTRD HIP 41HDX127ST FNSTRD HIP 42HDX127ST FNSTRD HIP 43HDX127ST FNSTRD HIP 44HDX127ST FNSTRD HIP 45HDX140ST FNSTRD HIP 46HDX140ST FNSTRD HIP 47HDX140ST FNSTRD HIP 48HDX140ST FNSTRD HIP 49HDX140ST FNSTRD HIP 50HDX152ST FNSTRD HIP 51HDX152ST FNSTRD HIP 52HDX152ST FNSTRD HIP 53HDX152ST FNSTRD HIP 54HDX152ST FNSTRD HIP 55HDX152S
| Date | Aug 26, 2014 (initiated Jul 22, 2014) |
| Status | Terminated |
| Reason | Zimmer is recalling sterile-packaged implants and instruments packaged in six different package configurations due to packaging design verification test failures. Specifically, multiple test samples from each of the six configurations failed simulated distribution and shipping testing when the devices either breached inner or outer package trays or seal integrity was lost. |
| Root cause | Package design/selection |
| Action | Zimmer sent an URGENT MEDICAL DEVICE RECALL letter dated July 2014 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. Please call the Customer Call Center at 1-800-348-2759 for questions or concerns involving this notice. |
| Quantity | 972,125 units |
| Distribution | Worldwide Distribution. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.