Recall Z-2653-2014 ยท IYE Accelerator, Linear, Medical

MITSUBISHI HEAVY INDUSTRIES, LTD., HIROSHIMA MACHINERY WORK

VERO Linear Accelerator System, Model Number MHI-TM2000; Common Name: Medical Linear Accelerator. MHI-TM2000 Linear Accelerator System is intended for radiation therapy of lesions, tumors and conditions anywhere in the body where radiation treatment is indicated.

DateSep 16, 2014 (initiated Jul 5, 2014)
StatusTerminated
ReasonSoftware Anomaly: If a user changes the calendar setting from Workday to Holiday or vice versa, that would erroneously change the status of treatment/fractions, which are completed and to become Treated (completed) status on the exact day when such change is made, to Untreated status. Furthermore, such treatment would be cloned and mistakenly added to the schedule as Untreated treatment
Root causeSoftware design
ActionA Preliminary Customer Information Letter (dated 7/4/14) was emailed to customers on 7/05/14 from Mitsubishi's sales and service representative in Germany. The letter informed customers of the potential safety issue associated with the VERO Linear Accelerator and the recommended actions to be taken. A formal URGENT FIELD SAFETY NOTICE Letter (dated 7/15/14) was emailed to customers on 8/07/14 from Mitsubishi's sales and service representative in Germany. The letter instructed customers to ensure that all potential users in the facility are made aware of the safety notification and the recom
Quantity3 units in use in the US
DistributionDistributed in the states of NY, FL, and TX.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K072047MHI-TM2000 LINEAR ACCELERATOR SYSTEM
Mitsubishi Heavy Industries, Ltd., Aug 29, 2007
K122450MHI-TM2000 LINEAR ACCELERATOR SYSTEM VERO IMAVIS
Mitsubishi Heavy Industries, Ltd., Oct 3, 2012