Recall Z-2255-2014 ยท JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer

Ortho Development Corporation

PS Tibial Insert, REF 163-1XXX Product Usage: UHMWPE component used in TKA procedures to serve as a tibial bearing surface for articulation with the femoral component

DateAug 17, 2014 (initiated Jul 11, 2014)
StatusTerminated
ReasonOrtho Development is recalling various Pivot Bipolar Cup, Patella, PS Tibial Insert, UC Tibila Insert, and CK Tibil Insert because the foil pouches did not properly seal, allowing air to seep back into the pouch after sealing which increases the risk that the material properties of the implant would be substantially degraded due to oxidation.
Root causePackaging process control
ActionOrtho Development notified all affected consignees/distributors of the recall via email on July 11, 2014 which included a list of the affected product (part numbers, lot numbers and quantities) under their control. An RMA number and return shipping label were also included with the email. Emails were followed up with phone calls to verify receipt of the notification.
Quantity11
DistributionWorldwide Distribution - US Nationwide and the country of Japan. No military/govt/VA consignees.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K994370ORTHO DEVELOPMENT BALANCED KNEE SYSTEM
Ortho Development Corp., Mar 24, 2000