ClearanceBrief
Recall Z-2183-2014 ยท NIG System, Test, Carbohydrate Antigen (Ca19-9), For Monitoring And Management Of Pancreatic Cancer

ORTHO-CLINICAL DIAGNOSTICS

VITROS Immunodiagnostic Products CA 19-9 Calibrators and Reagent Packs, Lot 1320 VITROS CA 19-9 Reagent Packs are used on VITROS ECi/ECiQ Immunodiagnostic Systems, VITROS 3600 Immunodiagnostic Systems and VITROS 5600 Integrated Systems for the quantitative measurement of 1116-NS-19-9 defined antigen in human serum and plasma.

DateAug 11, 2014 (initiated Jul 16, 2014)
StatusTerminated
ReasonCustomers may receive positively biased results when using VITROS Immunodiagnostic Products CA 19-9 Total Reagent Packs, Lot 1320 and VITROS CA 19-9 Calibrators, Lot 1320 on the VITROS ECi/ECiQ Immunodiagnostic Systems, the VITROS 3600 Immunodiagnostic System and the VITROS 5600 Integrated System. There is an unexpected observed change in CA 19-9 Calibrators, Lot 1320 calibrator values.
Root causeOther
ActionOrtho Clinical Diagnostics sent an Urgent Product Correction Notification July 2014 to all affected customers. The Notice identified the product, the problem, and the action to be taken by the customer. Customers were instructed to Immediately discontinue using and discard Lots 1320 of VITROS CA 19-9 Calibrators and Reagent Packs. Review previously reported VITROS 25-OH Vitamin D Total results using Lot 1320. Review quality control results used to verify calibration events. Results lower than the target mean may indicate potentially biased assay results. Discuss any concerns regarding previ
Quantity475
DistributionWorldwide Distribution - USA (nationwide) and Internationally to France and Japan
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.