Recall Z-2263-2014 ยท GEX Powered Laser Surgical Instrument

Biolase Technology Inc

WaterLase iPlus Dental Laser System, Part No. 7200854-XX Software Upgrade Kit, Part No. 6201463.

DateAug 18, 2014 (initiated Jun 12, 2014)
StatusTerminated
ReasonBiolase is recalling the WaterLase iPlus Dental Laser System because of a discrepancy present in the user interface software (version 1.7). If an error occurs in the procedure screen, the air setting will change to 1% and OFF.
Root causeUnder Investigation by firm
ActionAn Urgent Medical Device Correction letter dated 6/12/2014 was sent to all customers who purchased the WaterLase iPlus Dental Laser System. The letter informs the customers of the problems identified and the actions to be taken. Customers are informed that a Field Service Engineer can install the software update. Customers with are instructed to contact Customer Service at (800) 321-6717.
Quantity117 units
DistributionNationwide in US and worldwide to: Canada, UK, Russia, Korea, Hong Kong, Libya, Iran, Iraq, Kuwait, Singapore, Armenia, Egypt, Pakistan, Sweden, New Zealand, Poland, Germany, Italy, UAE.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K101658WATERLASE MD TURBO PLUS MODEL 7200XXX
Biolase Technology, Inc., Aug 11, 2010