ClearanceBrief
Recall Z-2296-2014 ยท JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer

Zimmer, Inc.

Persona (TM) The Personalized Knee System PS Tibial Articular Surface Provisional Top Nonsterile TASP PS L 3-5 CD TOP TASP PS L 3-5 CD TOP TASP PS L 3-5 CD TOP TASP PS L 3-5 CD TOP TASP PS L 3-5 CD TOP TASP PS L 3-5 CD TOP TASP PS L 3-5 CD TOP TASP PS L 3-5 CD TOP TASP PS L 3-5 CD TOP TASP PS L 3-5 CD TOP TASP PS L 3-5 CD TOP TASP PS L 3-5 CD TOP TASP PS L 3-5 CD TOP TASP PS L 3-5 CD TOP TASP PS

DateAug 21, 2014 (initiated Jun 10, 2013)
StatusTerminated
ReasonZimmer had received complaints that reported breakage of the Persona Knee Trial Articular Surface Provisionals (TASPs). This is a retrospective recall which covers Zimmer's corrective actions taken earlier this year concerning the TASPs.
Root causeDevice Design
ActionOn February 10, 2014, Zimmer sent letter to surgeons who had filed complaints regarding the affected TASP fractures. The letter provided the surgeons with Zimmer's investigational findings of the issue. Surgeons were directed to send questions or any additional informtion to Zimmer.PER@Zimmer.com. The Zimmer Weekly Wrap-Up, dated February 14, 2014, was issued electronically to sales force. They were notified to engage in mandatory training to improve the usage of the TASPs and prevent potential breakage. Zimmer has taken additional corrective measures by initiating another recall in July 2
Quantity70,986 distribution events
DistributionWorldwide Distribution-USA (nationwide) including DC and Puerto Rico and the states of AK, AL, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, MA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, SD, TN, TX, UT, VA, WA, WI, WY, and the countries of Australia, Canada, Germ
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K113369ZIMMER PERSONA KNEE SYSTEM
Zimmer, Inc., Mar 27, 2012