Recall Z-2171-2014 ยท NAY System, Surgical, Computer Controlled Instrument

Intuitive Surgical, Inc.

EndoWrist Stapler System User Manual Addendum and Wall Chart; The EndoWrist Stapler 45 System and EndoWrist Stapler 45 Reloads are intended to be used with the da Vinci Si Surgical System (Model IS3000) for resection, transection and/or creation of anastomoses in General, Gynecologic and Urologic surgery. The device can be used with staple line or tissue buttressing material (natural or syntheti

DateAug 8, 2014 (initiated Jun 2, 2014)
StatusTerminated
ReasonChanges were made to the Stapler EndoWrist Stapler 45 System Instruments and Accessories User Manual Addendum (550986-03) and the EndoWrist Stapler 45 Wall Chart (PN 551524-01) in which the pH range specification for the enzymatic cleaner used throughout the cleaning process was modified from pH-neutral to pH-neutral to mild alkaline (mild alkaline is defined as pHd11). Mild Alkaline enzymatic
Root causeLabeling Change Control
ActionIntuitive Surgical representatives hand delivered corrected manuals to the two affected sites..
Quantity2
DistributionDistribution to Florida and South Carolina
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K113706ENDOWRIST STAPLER SYSTEM
Intuitive Surgical, Inc., Oct 17, 2012