Recall Z-2156-2014 ยท LFL Instrument, Ultrasonic Surgical

Stryker Instruments Div. of Stryker Corporation

Sonopet Ultrasonic Surgical System Console, Ultrasonic Aspirator, Stryker Medtech.

DateAug 7, 2014 (initiated Jun 24, 2014)
StatusTerminated
ReasonThe firm received complaints regarding insufficient irrigation to the tip of the Sonopet system. Lack of irrigation can potentially lead to excessive heat at the tip.
Root causeLabeling design
ActionThe firm sent Medical Device Upgrade Notification letters dated July 11, 2014 to their customers.
Quantity2,460 units
DistributionNationwide Distribution-including DC and the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, VT, WA, WI & WV.