Recall Z-2209-2014 ยท LGX Table, Examination, Medical, Powered
Medical Positioning Inc
Medical Positioning Rapid Response HUT table, Powered Medical Examination table, Model # 1019. Product Usage: Used to conduct syncope examinations.
| Date | Aug 13, 2014 (initiated Jul 10, 2014) |
| Status | Terminated |
| Reason | Excessive weight on the table will cause tables with the affected component to drop into the Trendelenburg position more rapidly than normal. The product was manufactured with a component that can cause the device to malfunction. |
| Root cause | Component design/selection |
| Action | Medical Positioning Inc. sent Potential Product Hazard Bulletin letter dated July 8, 2014 to affected customers, via USPS certified return receipt mail. The letter identified the affected product, problem and actions to be taken. Customers were informed a Product Support Specialist will contact them to make arrangements for a Field Service Technician to perform a test designed to identify if the table component affects the table mechanics. If the table is identified as having the affected component, the component will be replaced. |
| Quantity | 2 beds |
| Distribution | Worldwide Distribution - US Nationwide in the states of CA, GA, FL, IL, IN, KS, KY, MD, MN, MO, MS, NC, NY, OH, TN, TX, VA and WI. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.