Recall Z-2194-2014 ยท MDP Separator For Therapeutic Purposes, Membrane Automated Blood Cell/Plasma
Gambro Renal Products, Incorporated
Prisma TPE 2000 set. Product number 109672. Each set is packaged in a plastic pouch, 4 pouches per carton box
| Date | Aug 13, 2014 (initiated Jul 18, 2014) |
| Status | Terminated |
| Reason | Gambro initiated a field action on the Prisma and Prismaflex sets due to breakages of connectors resulting in leakage of liquid or blood. |
| Root cause | Device Design |
| Action | Gambro sent an Urgent Field Safety Notice dated July 18, 2014 to all end-users. A Customer Reply Form was attached to the notice for customers to complete and return. |
| Quantity | 401,451 |
| Distribution | Worldwide Distribution. |
| FDA record | Open on fda.gov |