ClearanceBrief
Recall Z-2193-2014 ยท MDP Separator For Therapeutic Purposes, Membrane Automated Blood Cell/Plasma

Gambro Renal Products, Incorporated

Prismaflex TPE 2000 set, Product Number 1071441. Each set is packaged in a plastic pouch, 4 pouches per carton box

DateAug 13, 2014 (initiated Jul 18, 2014)
StatusTerminated
ReasonGambro initiated a field action on the Prisma and Prismaflex sets due to breakages of connectors resulting in leakage of liquid or blood.
Root causeDevice Design
ActionGambro sent an Urgent Field Safety Notice dated July 18, 2014 to all end-users. A Customer Reply Form was attached to the notice for customers to complete and return.
Quantity401,451
DistributionWorldwide Distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.