Recall Z-2292-2014 ยท JAA System, X-Ray, Fluoroscopic, Image-Intensified

Siemens Medical Solutions USA, Inc

Siemens Luminos dRF, Ysio or Uroskop Omnia systems. The Axiom Luminos dRF is intended to be used as a universal diagnostic imaging system for radiographic and fluoroscopic studies.

DateSep 10, 2014 (initiated Jul 7, 2014)
StatusTerminated
ReasonIt was discovered that during a RAD examination using Siemens Luminos dRF, Ysio or Uroskop Omnia systems with software version VB10C to VB10F and automatic exposure control, a highly unlikely malfunction can result in longer radiation exposure than required. This may also result in an overexposed image that is not of diagnostic quality. As a result the taken examination needs to be repeated. Thi
Root causeDevice Design
ActionSiemens had informed all affected customers about this issue via Customer Safety Advisory Notice XP023/13/S (dated June 3, 2013) reported under C&R # 2240869-05/16/13-0013-C. A corresponding software update to fix the issue was originally planned to be rolled out by May 2014. The originally scheduled roll out date could not be met and to ensure that this new software version works accurately Siemens incorporated a comprehensive monitoring phase. Siemens has now scheduled the roll out for the software version VB10J to be completed by end of October 2014. A letter, dated July 7, 2014, with im
Quantity192
DistributionNationwide Distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K062623AXIOM LUMINOS DRF
Siemens Medical Solutions USA, Inc., Aug 22, 2007
K081722YSIO
Siemens Medical Solutions USA, Inc., Aug 22, 2008
K101491UROSKOP OMNIA
Siemens Medical Solutions USA, Inc., Jun 30, 2010