Recall Z-2132-2014 ยท HQF Laser, Ophthalmic

Nidek Inc

MC-500 Multicolor Laser Photocoagulator; Distributed by NIDEK Inc. Manufactured by NIDEK CO., Ltd, Japan. For use in ophthalmic and retinal procedures.

DateJul 30, 2014 (initiated Jun 9, 2014)
StatusTerminated
ReasonLaser aperture label was not affixed to device prior to shipment in the US.
Root causeError in labeling
ActionNidek sent a Recall/Field Correction letter dated June 9, 2014, to all affected customers. The letter advised customers of the issues and informed customers that Nidek Representatives will visit each location to affix the proper labeling to the device. Customers with questions were instructed to contact Nidek Customer Service at 1-800-223-9044 M-F 8 am to 5 pm PST. For questions regarding this recall call 510-353-7785.
Quantity29 units (26 already re-labeled)
DistributionNationwide Distribution including IL NE NY, NH, CD, MI, TX, CA, PA, FL, KS, UT, MN, WI,. MO. and OR.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K110228NIDEK MULTICOLOR LASER PHOTOCOAGULATOR
Nidek Incorporated, Feb 23, 2011