ClearanceBrief
Recall Z-2110-2014 ยท LLZ System, Image Processing, Radiological

Philips Ultrasound, Inc.

QLAB 10.1 Core Module PN 453561728001. QLAB Quantification Software is a software application package designed to view and quantify image data acquired on Philips Medical Systems ultrasound products.

DateJul 23, 2014 (initiated Jul 9, 2014)
StatusTerminated
ReasonThe QLAB a2DQ and aCMQ features have a defect that may use incorrect measurement values under specific conditions.
Root causeSoftware change control
ActionPhilips Medical Systems notified customers of the voluntary recall by sending a certified URGENT-Medical Device Correction letter, dated 2014 JUL 09, to all US and Canadian consignees. Philips representatives will distribute the letter internationally. Follow-up letters to wholesale customers will be provided and wholesalers are requested to forward copies of recall letters and response forms to their customers. Philips Medical Systems will also have Customer Reply Cards returned for tracking purposes. Customers are directed to call local Philips representative or Philips Customer Service
Quantity59 units total (12 units in the US and 47 units outside the US)
DistributionDistributed in the states of CA, IA, KY, MD, NY, PA, UT, VA, WA and WI. Also distributed in Australia, Belgium, Canada, India, Italy, Japan, South Korea, Netherlands, New Zealand, Spain, Sweden, and Switzerland.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K152165S9 Portable Digital Color Doppler Ultrasound System
Sonoscape Medical Corp., Aug 26, 2015