Recall Z-2145-2014 ยท HET Laparoscope, Gynecologic (And Accessories)

Market-Tiers Inc

Instructions for Use for the blue endo MOREsolution Electromechanical Laparoscopic Morcellator System, Model 55-000011, consisting of the control unit, hand-grip, flexible shaft, trocars/obturators, sleeves, valves, and blades.

DateAug 2, 2014 (initiated Jul 16, 2014)
StatusTerminated
ReasonSome pages of the Instructions for Use were omitted, including pages regarding the Indications for Use, Warning, and Contraindications.
Root causeLabeling Change Control
ActionThe recalling firm issued a letter dated 7/16/14 via Federal Express requesting the customer destroy the incorrect IFU and replace it with the corrected version.
Quantity100 units
DistributionNationwide Distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K091010TROKAMED MORCELLATOR
Trokamed GmbH, Mar 1, 2011