Recall Z-2123-2014 ยท GEX Powered Laser Surgical Instrument

Ra Medical Systems Inc

PHAROS Excimer Laser, Model No. EX-308 UVB phototherapy for the treatment of psoriasis, vitiligo, atopic dermatitis and leukoderma

DateJul 28, 2014 (initiated Jun 30, 2014)
StatusTerminated
ReasonRA Medical Systems is recalling the PHAROS EX-308 Excimer Laser because there is a possibility that the laser may calibrate with the iris closed.
Root causeError in labeling
ActionRa Medical Systems sent a customer letter dated June 30, 2014, to who purchased the PHAROS EX-308 Excimer Laser Phototherapy System Hand Piece. The letter informed the customers of the problems identified and the actions to be taken. Customers were also provided with the Instructions for Use (IFU) for the PHAROS EX-308 Excimer Laser Phototherapy System Hand Piece and a replacement iris indication label for the handpiece.
Quantity435 units
DistributionWorldwide Distribution - USA (Nationwide) and Internationally to: Korea, Kuwait, and Saudi Arabia.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K062963PHAROS EXCIMER LASER SYSTEM, MODEL EX-308
Ra Medical Systems,Inc, Apr 3, 2007