Recall Z-2167-2014 ยท JOP Transducer, Ultrasonic

Biosound Esaote, Inc.

Esaote PA230E Transducer A portable and cart base ultrasound system for general and cardiovascular imaging

DateAug 8, 2014 (initiated Jun 26, 2014)
StatusTerminated
ReasonUsing the PA230E transducer with a user-defined setting with the output power to 100% (Max) in 2D-TEI mode focusing at maximum depth, could cause the transducer surface temperature to exceed the maximum allowable limit.
Root causeSoftware design
ActionOn 06/26/2014, Esaote North America sent an advisory letter via FEDEX (with signature required) to each customer or consignee affected by this voluntary recall. The packet contained the advisory letter, including actions to be taken by the user, information about a pending software upgrade, and a business reply card to confirm they received the letter and have the equipment noted on the advisory letter. Customers with questions were instructed to call 800-428-7378. For questions regarding this recall call 1-317-813-6005
Quantity1,301 units
DistributionNationwide-Product was shipped to the following states: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, SC, SD, TN, TX, UT, VA, WA, WI & WV. International**Product was also shipped to consignees i
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K040596ULTRASOUND IMAGING SYSTEM, MODEL 7300
Esaote, S.p.A., Mar 31, 2004
K0503267350 ULTRASOUND IMAGING SYSTEM, MODEL 7350
Esaote, S.p.A., Feb 28, 2005
K110688MYLABGOLD
Esaote, Jun 2, 2011