Recall Z-2153-2014 ยท N/A

PerkinElmer Health Sciences, Inc.

Panthera-Puncher 9 . Panthera-Puncher 9 is a sample transfer instrument for professional use. The sample, a dried blood disk, is extracted from a larger area of the sample (dried blood spot) by punching the disk of the specified size to a sample collection vessel, e.g. a 96-well microtiter plate. The identification of the sample punched to a specific well of the plate is stored and carried on t

DateAug 6, 2014 (initiated Jul 1, 2014)
StatusTerminated
ReasonRisk of incorrect results from samples prepared with Panthera-Puncher 9
Root causeProcess design
ActionPerkin Elmer sent an Urgent Medical Device Recall letter dated July 1, 2014, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The Recall Letter delivered to the end customer by email facilitated through our Perkin Elmer Technical Support group located in Akron Ohio. Technical Support will be asked to make sure the customers receive the Recall Letter and implement the recommended precautions: "Removing the plates from the Panthera-Puncher 9 when not actively punching and processing samples in the instrument. "Opening
Quantity144 units
DistributionWorldwide Distribution - USA (nationwide) and the countries of : Canada, Australia, Brazil, China, Denmark, Great Britain, Italy, France, Mexico, Qatar and Korea.