Recall Z-2149-2014 ยท NBC Test, Natriuretic Peptide

Siemens Healthcare Diagnostics, Inc

SIEMENS ADVIA Centaur CP Immunoassay System BNP kit; SMN 10309044 - 100 test For in vitro diagnostic use in the quantitative determination of B-type Natriuretic Peptide (BNP) in human plasma using the ADVIA Centaur CP System.

DateJun 26, 2015 (initiated Jun 23, 2014)
StatusTerminated
ReasonComplaints of failed calibration.
Root causeOther
ActionSiemens sent an Urgent Medical Device Correction (UMDC) letter to all affected Siemens Healthcare Diagnostics customers in the United States on June 23, 2014 via Federal Express. An Urgent Field Safety Notice (UFSN) was sent out for distribution to all customers outside the United States on June 23, 2014 via hard copy, email, and/or fax per each country's local regulations. These notices inform the customer of the potential for low calibrator deviation failures when using ADVIA Centaur Systems BNP kit lots ending in 172 on the ADVIA Centaur CP due to an error on the master curve card. Siemens
Quantity1,191 kits
DistributionWorldwide Distribution - USA (nationwide including Puerto Rico) and Internationally to Canada, Albania, Argentina, Australia, Bosnia Herzeg., Brazil, Chile, Colombia, Egypt, Georgia, Hong Kong, India, Indonesia, Iran, Iraq, Israel, Japan, Kazakhstan, Lebanon, Malaysia, New Caledonia, Oman, P.R. Chin
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K031038ADVIA CENTAUR B-TYPE NATRIURETIC PEPTIDE (BNP) ASSAY
Bayer Healthcare, LLC, Jun 23, 2003