Recall Z-2567-2014 ยท IYE Accelerator, Linear, Medical

Varian Medical Systems, Inc.

C-Series High Energy Linear Accelerator - Models: Novalis Tx, Trilogy, Trilogy Tx, Clinac iX, Clinac 2100 C/D, Clinac 2300C/D, Clinac 21EX, Clinac 23EX with versions 7, 8, 9 and 6 MV configuration and maximum dose rate of at least 6000 MU/min only. Varian High Energy Linear Accelerator is intended to provide sterotactic radiosurgery and precision radio-therapy for lesions, tumors and conditions a

DateSep 13, 2014 (initiated Jun 17, 2014)
StatusTerminated
ReasonTrend in reports of unexpected decrease in beam output in C-series High Energy Linear Accelerators for 6MV photon treatment mode.
Root causeRadiation Control for Health and Safety Act
ActionUrgent Medical Device correction letters were sent on June 16, 2017 to all affected users by certified mail.
Quantity3416
DistributionWorldwide Distribution-USA (nationwide) and the countries of Algeria, Argentina, Australia, Austria, Azerbaijan, Bangladesh, Byelarus, Belgium, Bosniak, Brazil;, Bulgaria, Burma, Canada, Chile, China, Columbia, Costa Rica, Czech Republic, Denmark, Dominican Republic, Ecuador, Estonia, Finland, Fran
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.