Recall Z-2118-2014 ยท HAW Neurological Stereotaxic Instrument

Medtronic Neuromodulation

Medtronic Nexdrive Micropositioning Drive. Models Ml-1000 and Ml-2000. For use in conjunction with the Medtronic Nexframe Stereotactic System for the precise positioning of microelectrodes and implantable leads. A stereotactic guidance system used in conjunction with Medtronic StealthStation Navigation Systems-image-guided surgery (IGS) systems-for Deep Brain Stimulation (DBS) procedures.

DateJul 25, 2014 (initiated Jun 11, 2014)
StatusTerminated
ReasonPotential for misalignment of the Z-stage scale. Using one of these devices for a procedure could result in the microelectrode being inserted to an incorrect target depth.
Root causeNonconforming Material/Component
ActionMedtronic sent an Urgent Medical Device Recall letter dated June 2014 on June 11, 2014, to all affected customers. The letter described the problem and the product involved in the recall. The letter described the "Action Requested" which included to stop using and return the recalled device and to complete and return the Reply Form. Customers with questions were instructed to contact Medtronic Neuromodulation Technical Services at 1-800-707-0933. EXPANDED 10/02/2014: Medtronic sent an Urgent Medical Device Recall letter beginning 10/02/2014. The letter explained that this letter is an up
Quantity300 EXPANDED 10/02/2014 130 devices
DistributionWorldwide Distribution - USA including CA, CO, FL, IL, KS, KY, LA, MN, MT, NC, NE, OR, TX, VA, and UT, and Internationally to Canada, Spain, and United Kingdom.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K022100NAVIGUS MANUAL DISPOSABLE MICRODRIVE WITH DIGITAL DISPLAY, MODELS MI-1000 & MI-1010
Image-Guided Neurologics, Inc., Sep 25, 2002