Recall Z-2116-2014 ยท FDS Gastroscope And Accessories, Flexible/Rigid

Pentax Medical Company

EG-3630U Ultrasound Video Gastroscope EG-3830UT Ultrasound Video Gastroscope EG-3870UTK Ultrasound Video Gastroscope FG-36UX Ultrasound Fiber Gastroscope Used to provide optical and sonographic visualization of and therapeutic access to the Upper Gastrointestinal Tract.

DateJul 24, 2014 (initiated Jun 12, 2014)
StatusTerminated
ReasonAspiration needles used in combination with Ultrasound Gastroscopes can derail from the accessory elevator at the distal end of the endoscope. This may result in patient injury if the Instruction for Use is not followed carefully by the user.
Root causeDevice Design
ActionPentax Medical sent an Urgent Medical Device Field Safety Corrective Action dated June 12, 2014 to affected customers. The letter identified the affected product, problem and actions to be taken. Pentax also posted their notification on their website. For questions regarding this action please feel free to contact Paul Silva, Recall Coordinator at Tel: 800-431-5880 ext 2064 (8:30am 5:00pm Mon-Fri, EST) Fax: 201-799-4063 Email : paul.silva@pentaxmedical.com
Quantity531 units
DistributionUS Nationwide Distribution