Recall Z-2085-2014 ยท MHX Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)

GE Healthcare, LLC

GE Healthcare, CARESCAPE Monitor B850, B650 or B450. Patient Monitor.

DateJul 21, 2014 (initiated Jun 13, 2014)
StatusTerminated
ReasonPartial loss of monitoring with the CARESCAPE Monitor B850, B650 or B450. When connecting the PDM (Patient Data Module) to the CARESCAPE B850, B650 or B450 monitor with software version 2 the PDM may not start communicating to the monitor and loss of the PDM parameters could occur.
Root causeSoftware design
ActionConsignees were sent on 6/13/2014 a GE Healthcare "Urgent Medical Device Correction" letter #36103, 36104 and 36105 dated June 6, 2014. The letter was addressed to Healthcare Administrator / Risk Manger, Chief of Nursing and Director of Biomedical Engineering. The letter described the Safety Issue, Safety Instructions, Affected Products Details, Product Correction and Contact Information. Consignees were advised to contact Technical Support at 800-558-7044 in regard to questions on this recall.
Quantity41,992 (20,509 US; 21,483 OUS)
DistributionWorldwide Distribution: US nationwide, UNITED KINGDOM, UNITED ARAB EMIRATES, TURKEY, TAIWAN, SWITZERLAND, SWEDEN, SPAIN, SOUTH AFRICA, SLOVENIA, SLOVAKIA, SINGAPORE, SERBIA, SAUDI ARABIA, RUSSIAN FEDERATION, ROMANIA, REUNION, QATAR, PORTUGAL, POLAND, PHILIPPINEs, PARAGUAY, PANAMA, NORWAY, NEW ZEALAN
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K092027CARESCAPE MONITOR B850
Ge Medical Systems Information Technologies, Aug 28, 2009
K131223CARESCAPE MONITOR B650
GE Healthcare Finland Oy, Aug 28, 2013
K131414CARESCAPE MONITOR B850
GE Healthcare Finland Oy, Aug 28, 2013
K132523MYOPERFUSION
Toshibamedical Systems Corporation, Dec 12, 2013
K192239WavelinQ Plus EndoAVF System
C.R. Bard, Inc., Oct 17, 2019