Recall Z-1996-2014 ยท FPA Set, Administration, Intravascular

RGI Medical Manufacturing, Inc.

Henora REF: 70-4026-09V1 300psi I.V. Catheter Extension Set Product Usage: Catheter extension set for high pressure procedures.

DateJul 7, 2014 (initiated Mar 28, 2014)
StatusTerminated
ReasonProblem with low adhesion of the tubing to luer connection which may lead to a separation of the extension set while in use.
Root causeProcess control
ActionRGI Medical Manufacturing, Inc. phoned their direct account: Baxter Healthcare on February 28, 2014 and sent an Urgent: Device Recall letter dated March 21, 2014. The letter identified the affected product, problem and the actions to be taken. Contact 800-747-6870 to obtain instruction for returning the product to RGI
Quantity19,200 units.
DistributionUS Nationwide Distribution in the states of AL, CA, DE, GA, IL, KY, MA, MD, ME, MI, MN, NJ, NM, NV, NY, OH, TN, TX, WA, and WI.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K001599ACTI-FLEX IV ADMINISTRATION SET
R-Group Intl., Sep 18, 2000