ClearanceBrief
Recall Z-1988-2014 ยท KPR System, X-Ray, Stationary

GE Healthcare, LLC

GE Healthcare, Proteus XR/a. Proteus XR/a Intended Use: Is intended for use in generating radiographic images of human anatomy in all general purpose diagnostic procedures. This device is not intended for mammographic applications.

DateJul 2, 2014 (initiated Jun 4, 2014)
StatusTerminated
ReasonGE Healthcare has become aware of a potential safety issue due to a collimator installation error during a service maintenance activity involving the Proteus XR/a and Revolution XR/d X-ray imaging systems. . There was a reported incident of a collimator fall which caused a serious patient injury.
Root causeEquipment maintenance
ActionGE Healthcare sent an Urgent Medical Device Correction letter dated June 4, 2014, to all affected consignees. The letter described the Safety Issue, Safety Instructions, Affected Product Details, Product Correction and Contact Information. For questions in regard to this recall the consignees in the United States were advised to call 800-437-1171. For other countries contact your local GE Healthcare Service Representative. For questions regarding this recall call 800-437-1171.
Quantity3460
DistributionWorldwide Distribution - USA including PR. DC, GUAM, all states except VT. Internationally to the following countries: ALBANIA, ALGERIA, ARGENTINA, AUSTRALIA, AUSTRIA, BAHRAIN, BENIN, BOSNIA AND HERZEGOVINA, BOTSWANA, BRAZIL, BULGARIA, BURKINA FASO, CANADA, CHILE, CHINA, COLOMBIA, CZECH REPUBLIC, D
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K993090SILHOUETTE FC
Ge Medical Systems, Inc., Oct 1, 1999