Recall Z-1811-2014 ยท HXX Screwdriver

Ebi, Llc

Biomet Spine, Cypher MIS Screw Inserter, Catalog number 14-501669, instrument for Spinal Screw Fixation System.

DateJun 17, 2014 (initiated May 15, 2014)
StatusTerminated
ReasonThe Cypher MIS Screw Inserter may exhibit an increased rate of instrument tip failure.
Root causeNonconforming Material/Component
ActionBiomet sent an "Urgent Medical Device Recall Notice"/Fax Back Response Form dated May 15, 2014, via Fed Ex to all affected customers. The notice identified the product, the problem, and the action to be taken by the customer. Customers were instructed to inspect their inventory for the affected product and if located to remove the product from circulation; follow instructions on the attached Fax Back Response Form prior to return of the product; and return the product using the Fed Ex label included with the notification. Customers were instructed, if if they further distributed the produc
Quantity32 units (30 domestic and 2 international)
DistributionWorldwide Distribution - USA (nationwide) and the country The Netherlands.
FDA recordOpen on fda.gov