Recall Z-2698-2014 ยท MSD Catheter, Hemodialysis, Implanted
Argon Medical Devices, Inc
UltraStream Peel Away Kit; Model Numbers 32401524 (24 cm), 32401528 (28cm), 32401532 (32 cm), and 32401536 (36 cm) and 32402540 (40 cm) The Rex Medical UltraStream Chronic Hemodialysis Catheter is a chronic hemodialysis catheter that can achieve high flow rates at low arterial pressure.
| Date | Sep 24, 2014 (initiated Jun 5, 2014) |
| Status | Terminated |
| Reason | UltraStream Chronic Hemodialysis Catheter may develop cracks on the red (arterial) luer during attachment or manipulation. |
| Root cause | Component design/selection |
| Quantity | 3,783 units |
| Distribution | Nationwide Distribution. |
| FDA record | Open on fda.gov |