Recall Z-2174-2014 ยท IYE Accelerator, Linear, Medical

Siemens Medical Solutions USA, Inc

Siemens Linear Accelerators of type ARTISTE, ONCOR, and PRIMUS with Automatic Sequenced Cancer Treatment Delivery Option.

DateAug 10, 2014 (initiated May 28, 2014)
StatusTerminated
ReasonA safety risk exists with automatically sequenced treatment technique using the SIMTEC Auto Field Option where automatic movements of the gantry and the tabletop during an auto-sequenced treatment have the potential to lead to a collision with the patient.
Root causeUse error
ActionAn Urgent Medical Device Correction Field Safety Notice Letter was sent to end users on/about June 2014 which described the product, problem, and actions to be taken to avoid encountering the safety risk to patients and users.
Quantity160
DistributionNationwide and ROW, including Canada and Mexico.
FDA recordOpen on fda.gov