Recall Z-1961-2014 ยท NKB Thoracolumbosacral Pedicle Screw System

Integra LifeSciences Corp.

MIS Rod 2, a component of the NewPort System. Model No: 60-0545 (45mm); 60-0550 (50mm); 60-0555 (55mm); 60-0560 (60mm); 60-0575 (75mm); 60-0580 (80mm). The NewPort MIS Rod 2 is a 5.5 diameter titanium spinal rod used to immobilize adjacent NewPort pedicle screws

DateJul 1, 2014 (initiated Jun 9, 2014)
StatusTerminated
ReasonIntegra Lifesciences is recalling NewPort System MIS Rod, 2 because it may be up to 30 degrees offset from specification requirements causing misalignment with the inserter. This could potentially result in the lordosis of the rod not aligning with the handle when the inserter is used in accordance with the current NewPort MIS System Surgical Technique.
Root causeUnder Investigation by firm
ActionA customer notification letter was sent on 6/9/14 to their customers and distributors who purchased the Integra NewPort System MIS Rod, 2 to inform them there may be a potential that the hex on some NewPort System MIS Rod, 2 are up to 30 degrees offset from specification requirements causing misalignment with the inserter. Customers are instructed to complete and return the customer response form and return by email or fax as indicated on the form. Customers with questions are instructed to contact Customer Service at (866) 942-8698.
Quantity94 units
DistributionWorldwide Distribution. US including states of CA, FL, IL, MD, MO, TN, TX, and countries of Australia, Czech Republic and Spain.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K083089NEWPORT SPINAL SYSTEM
Seaspine, Inc., Jan 13, 2009