Recall Z-1977-2014 ยท OYE Flow Cytometric Reagents And Accessories.

BD Biosciences, Systems & Reagents

BD FACS Canto and associated Fluidics Cart Product Usage: Flow cytometer for use in In vitro Diagnostics.

DateJul 2, 2014 (initiated Jun 6, 2014)
StatusTerminated
ReasonLaser cooling fans and internal fan of the associated fluidics cart are not operational.
Root causeNonconforming Material/Component
ActionBD sent an Important Product Recall notification letters dated June 2014 to affected US customers by trackable mail. International affiliates were notified on June 6, 2014. The letter identified the affected product, problem and the actions to be taken. A Field Change notification has been established and a BD field Service Engineer will coordinate with each account to perform the necessary corrections. Questions or concerns may be directed to BD Customer Support at BD Biosciences, prompt 2. Out of US customers are to contact local BD rep with questions.
Quantity35 installed devices
DistributionWorldwide Distribution - US Nationwide in the states of CA, OH, RI, GA, FL, MD, NY, WV, NJ, Ma, CT, IA, NM, MN and the countries of China and Belgium.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.